SOP-019: Operation of an Inverted Microscope
Purpose
To ensure the safe and standardized operation of an inverted microscope for routine examination of mammalian cell cultures.
Scope
This SOP applies to the routine use of inverted bright-field or phase-contrast microscopes for monitoring mammalian cell cultures.
Responsibilities
The operator is responsible for:
- Verifying microscope functionality before use.
- Handling objectives and optical components correctly.
- Preventing contamination of the microscope.
- Cleaning the instrument after use.
Required Materials
Reagents
- Lens cleaning solution (if required)
- Lens tissue
Equipment
- Inverted microscope
- Cell culture vessels
- Laboratory wipes
- Personal protective equipment (PPE)
Procedure
1. Pre-Use Inspection
Before use, verify that:
- The microscope is clean.
- Objectives and eyepieces are free of dust or residue.
- Illumination is functioning correctly.
- The stage moves smoothly.
2. Position the Sample
Place the culture vessel securely on the microscope stage.
Ensure that the vessel is stable before focusing.
3. Focus the Sample
Begin observation using the lowest magnification objective.
Bring the cells into focus using the coarse focus control, followed by the fine focus control.
Increase magnification only after the specimen has been properly focused.
4. Examine the Culture
Assess the culture for:
- Cell morphology
- Cell density
- Cell attachment (adherent cultures)
- Cell aggregation (suspension cultures)
- Evidence of contamination
Record observations when required.
5. Complete the Procedure
Remove the culture vessel.
Return the lowest magnification objective to the viewing position.
Switch off the illumination.
Clean the stage if contamination is present.
Cover the microscope after use.
Acceptance Criteria
| Parameter | Acceptance Criterion |
|---|---|
| Image quality | Clear and well focused |
| Optical components | Clean and undamaged |
| Stage movement | Smooth and unrestricted |
| Illumination | Uniform and stable |
Critical Control Points
- Begin focusing with the lowest magnification objective.
- Avoid contact between the objective lens and the culture vessel.
- Handle focusing controls gently.
- Keep optical components clean.
- Return the microscope to its storage position after use.
Documentation
Record the following when applicable:
- Date
- Cell line
- Observations
- Magnification used (if required)
- Operator initials
Quality Control
Verify:
- Image quality is satisfactory.
- Objectives are clean.
- Illumination functions normally.
- Stage movement is unrestricted.
- Microscope is cleaned after use.
Troubleshooting
| Observation | Possible Cause | Corrective Action |
|---|---|---|
| Blurred image | Dirty objective or incorrect focus | Clean the objective and refocus |
| Poor illumination | Lamp or illumination settings incorrect | Verify illumination settings and light source |
| Difficulty focusing | Incorrect objective selection or stage position | Return to the lowest magnification and refocus |
| Stage movement restricted | Mechanical obstruction | Remove obstructions and inspect the stage |
Safety Considerations
- Wear appropriate personal protective equipment (PPE).
- Handle culture vessels carefully to avoid spills.
- Clean any biological contamination immediately.
- Use only approved lens tissue for cleaning optical components.
- Disconnect the microscope from the power supply before maintenance.